AND UB-542 사용 설명서 - 페이지 7

{카테고리_이름} AND UB-542에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. AND UB-542 19 페이지. Wrist digital blood pressure monitor

AND UB-542 사용 설명서
Symbols that appear on the display (continued)
Symbols
The battery power indicator during measurement.
Full Battery
The battery is low when it blinks.
Low Battery
Unstable blood pressure due to movement during
measurement.
or
The systolic and diastolic values are within 10 mmHg of
each other.
The pressure value did not increase during inflation.
The cuff is not applied correctly.
The pulse is not detected correctly
PUL. DISPLAY ERROR
Blood pressure monitor internal error
Systolic blood pressure in mmHg
SYS.
Diastolic blood pressure in mmHg.
DIA.
Pulse per minute
PUL./min.
AM
Data taken between 4:00 and 9:59
PM
Data taken between 18:00 and 1:59
Symbols
Function / Meaning
English 6
Recommended Action
Replace all batteries with new ones when
the indicator blinks.
Take another measurement.
Remain still during measurement.
Apply the cuff correctly, and take another
measurement.
Remove the batteries and press the
button, and then install the batteries again.
If the error still appears, contact the dealer.
START