2.5 CALIBRATION VERIFICATION - UTILIZING EXTERNAL PRESSURE SOURCE
2.5 CALIBRATION VERIFICATION - UTILIZING EXTERNAL PRESSURE SOURCE
2.5 CALIBRATION VERIFICATION - UTILIZING EXTERNAL PRESSURE SOURCE
2.5 CALIBRATION VERIFICATION - UTILIZING EXTERNAL PRESSURE SOURCE
2.5 CALIBRATION VERIFICATION - UTILIZING EXTERNAL PRESSURE SOURCE
If available, an external pressure calibrator may be used to determine the current calibra-
tion of a sensor. Test procedure is as follows:
FIGURE 2-6 PRESSURE PUMP
FIGURE 2-6 PRESSURE PUMP
FIGURE 2-6 PRESSURE PUMP
FIGURE 2-6 PRESSURE PUMP
FIGURE 2-6 PRESSURE PUMP
HOOKUP
HOOKUP
HOOKUP
HOOKUP
HOOKUP
Digital Pressure Calibrator
(Suggested Unit)
Crystal Engineering
Phone: 800-444-1850
Model: 212-030PSI-G-HR
Pressure Pump
Snap fit Calibration Adapter
#73198-A3
P
9
AGE