4...20-mA signal
•
Filling level height for 4/20-mA signal
•
"Dry run" warning signal
•
"Failure" error signal
•
"Underdrive/overdrive" signal limit
•
"Underflow/overflow" error signal
•
Signal simulation (3.95...20.05 mA)
A list of the parameters settings in the continuous level sensor is included with the delivery. These de-
fault parameter values, and the parameter values changed by the user, can be printed out using the MPI-
200 programming adapter software.
When making the settings, note the help texts in the MPI software. These contain additional useful
information on changing the selected parameter.
6.1 Setup using the MPI-200 programming adapter
The MPI-200 programming adapter is connected to the NSL-F continuous level sensor via the ex-
ternal MPI-200-F adapter piece. It must be ensured that the NSL-F continuous level sensor is per-
manently connected to the supply voltage while the parameters are being set.
Connection of the programming adapter MPI-200
1: Connection for M12 connector
(external power supply)
2: USB port for connecting
to a PC incl. power
supply if not supplied externally
3: Connecting cable to
adapter for NSL-F
Signal flow during programming
Filling level measurement
•
Filling level zero point/offset
•
Filling level slope/gain
•
Damping/filter
•
Physical unit
Connection plug for MPI-200-F adapter as
an intermediate plug between the NSL-F
electronics and the MPI-200 connection 3
(see next figure).
1
3