Maintenance
5. Maintenance
14
The frequency of regular maintenance depends
on the degree of use, together with the
thickness of the test panels used –
• Heavy use – every 3 months
• Medium use – every 6 months
• Low use – every 12 months
1. Conduct a functional check to ensure that
the clamp plate is free to rotate, treat the
screw threads sparingly with light grease.
2. Turn the operating handles for complete
travel of the indenter, (nominally 20 mm).
Ensure the movement is smooth and even.
3. Inspect the hemispherical indenter for any
signs of wear, for example, fl ats on the
surface. Replace if necessary.
4. Check the magnifying glass condition. Clean
if necessary.
5. Inspect the batteries physical condition for
leakage in the magnifying glass. Remove and
replace if necessary.
6. Remove batteries if leakage occurs. Clean the
housing and fi t new batteries and re-check
function.
*Remove the batteries from the magnifi er if the byko-cup M is not
to be used for prolonged periods.