Placing the infusion / transfusion bag in the device
With the cover fully open, position the previously perforated bag on the plate. If the bag
model allows it, use the fixing hook on the plate. Otherwise, with your hand, hold
this pocket in place to prevent it from sliding forward when closing the system.
Remark : If the bag shape is longer than the plate, fold the bag end on itself in
order to meet a length ± equivalent to the plate length («Toothpaste tube »
technic).
Closing
Stand directly above the device with outstretched arms and press down on the cover.
Push down until the fastening clips are engaged audible by the "Click" sound.
The system is then closed and primed. This operation pressurises the bag.
Caution: In case only one fastening clip has been inserted, the plate will show an inclined
position. In this event, just press the not engaged side down in order to correctly position the
cover.
Positioning with respect to the patient
Ideally, the device should be positioned at the patient's midaxillary line level.
If placed at a higher level, the pressure shall be increased and conversely if it is placed at a lower
level (a 1 cm height difference is equivalent to ± 1 mbar).
Infusion / transfusion progress
The administration stage may be assessed by the triangle
drawn on the front surface of the mobile plate. The rising
action of the plate during administration will show the
different lines. The first line is wide and green in color. The
next two lines are red. The tip of the pyramid indicates that the
plate has reached its end point.
Instructions for use - Version 2