The Visulux software will allow for power measurement of the YAG laser. If a power meter is
available, the laser power can be measured with the YAG power monitor and the camera
removed from the output of the YAG laser. The camera is then installed on the laser, and the
Settings toolbar opened. Then open the Calibrate Power/Energy window. Either the laser
power as measured above can be entered, or the voltage as measured from the YAG power
monitor can be entered. Click on OK, and this power reading will appear in the data table. It is
important to note that this power reading will be valid only for the same value of exposure time
as when the calibration parameters were entered. It is highly unlikely that the relationship
between power readout and exposure time setting is linear, so if the exposure time is changed
the power should be recalibrated. Any changes in the neutral-density filter will also change this
value. The exposure time is being run at or below the minimum recommended value, and this
will result in considerable variation in the power readout with the camera.
If the above tests show that the SP2 YAG laser beam is in a satisfactory mode, no further
adjustments are needed. Upgrades to the Visulux software are planned in the future to allow for
continuous logging of the data, and this may be useful while collecting data with the SP2.
Figures 4-6 show measurements of a YAG laser that is in a higher-order mode. These can appear
as either a distortion of a single spot, or even multiple distinct spots. If both the X and Y axis are
in higher-order modes, multiple spots in both directions can appear.
Figure 4: The SP2 Laser Beam in a Higher-Order Mode, as viewed in the vBeam Control Window