8. After running the FlowCheck 9434031 beads, find another tube in the refrigerator
labeled "675 beads 763253." Open the "1A_675 Beads 763253" Acquisition Protocol
under FLOW. Be sure that the numbers on the tube match up with the numbers in the
Acquisition Protocol.
9. Run the sample as in 3-7. The HP X-CV values should be no more than 2.00.
10. Return the tubes to the same place you found them in the refrigerator.
Put the printouts in the FC500 Log Book vol. 2 binder sitting on top of the bookshelf on
the other side of the wall from the telephone.
If the volume in either tube gets too low, locate the appropriate small bottle of beads in
the refrigerator, right in front of where the tubes of beads are. The one with a blue dot on
top & a pink line around it is the 675 Bead bottle; the one with the blue line is the
FlowCheck 9434031 bottle. VORTEX the bottle for 10 seconds, then put enough drops
into the appropriate flow sample tube to bring up the volume. DO NOT TOUCH THE
TIP TO ANY SURFACE! This will contaminate the beads & make them un-useable (a
loss of several hundred dollars)