Check Calibration Status
Select Rgts/Cal from the menu bar
Select Print F10
Note: Calibration status can also be viewed on the software
Review the Cal Time Left and Cal Status
columns to identify reagents requiring
calibration
UniCel DxC, Customer, Daily Start Up Job Aid
Page 10 of 15
Rev. F (September 2013)
Beckman Coulter Professional Development