Ankom X59 교체 - 페이지 4

{카테고리_이름} Ankom X59에 대한 교체을 온라인으로 검색하거나 PDF를 다운로드하세요. Ankom X59 4 페이지. Analog level sensor and fiber optic cable with glass tip

Ankom X59 교체
9.
To remove Fiberoptic cable cut all cable ties that attach it to other wiring.
10.
Remove level sensor fitting from the back of the vessel using a 11/16 or adjustable wrench.
11. To install new fiberoptic and glass tip reverse steps 10-8 along with 5
12. Once you have installed the new X59 Analog level sensor module and Fiberoptic cable with glass tip. You are
now ready for testing
VERY IMPORTANT: As soon as there is pressure in the vessel apply a soap solution to all threads on the vessel fitting.
Make sure there are no leaks at the fitting or where the cable attaches to the fitting. Contact ANKOM with any
questions.
ANKOM Technology • 2052 O'Neil Road • Email: [email protected]• Phone: 315-986-8090 • Fax: 315-986-8091
Replacing X59 Analog level sensor
and Fiber optic cable with glass tip
XT15/XT15I
Service Procedure
X5S012
JAD 4/2/2019
pg. 4 of 4