Druck DPI 841 사용자 설명서 - 페이지 14

{카테고리_이름} Druck DPI 841에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Druck DPI 841 14 페이지. Frequency calibrator and frequency loop calibrator

7. To make sure that the calibration is correct, set up the
DPI 841/842 with these conditions:
Output Hz = 0 (For direct current output)
Output amplitude: Set to Unipolar (Table 5)
8. Supply the specified values (Table 13): Square Wave
then Sine Wave. Make sure that the error is in the
specified limits.
Table 13: Amplitude output error limits
Amplitude
Permitted
Volts (V)
DPI 841/842 error
(V)
0.1
0.05
1.0
0.10
10.0
0.10
15.0
0.15
24.0
0.24
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.

Specification data

All accuracy statements are for one year.
Specification - General
Languages
English [Default]
Operating
14 ... 122°F (-10 ... 50°C)
temperature
Storage
-4 ... 158°F (-20 ... 70°C)
temperature
Humidity
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
Shock/Vibration
BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
EMC
BS EN 61326-1:1998 + A2:2001
Safety
Electrical - BS EN 61010:2001; CE Marked
Size (L: W: H)
7.1 x 3.3 x 2.0 in
(180 x 85 x 50 mm)
Weight
14 oz (400 g)
Power supply
3 x AA alkaline batteries
Hz, pulses: ≈ 60 hours
Duration
mA: ≈ 35 hours
(Measure)
mA: ≈ 10 hours (24 V Source at 12 mA)
Hz, pulses: ≈ 20 hours
Duration
(Supply)
99 Washington Street, Melrose, MA 02176
12
- [EN] English
Test Equipment Depot - 800.517.8431
TestEquipmentDepot.com
Specification - Frequency
cpm = Pulses/minute, cph = Pulses/hour
Range (Measure):
0 ... 999.999 Hz
1 ... 50.0000 kHz
cpm: 0 ... 99999
cph: 0 ... 99999
Range (Supply):
0 ... 999.99 Hz
1 ... 50.000 kHz
cpm: 0 ... 99999
cph: 0 ... 99999
Temperature coefficient
14 ... 50°F, 86 ... 122°F
(-10 ... 10°C, 30 ... 50°C)
Waveform
(Sine, square, triangle)
Voltage input
Trigger level
Output amplitude
Connectors (A1 - Item 10)
Specification - Electrical connectors (A2)
Range (Measure)
Accuracy
Temperature coefficient
14 ... 50°F, 86 ... 122°F
(-10 ... 10°C, 30 ... 50°C)
Switch detection
Loop power output
®
HART
resistor
Connectors (A2)
Accuracy:
For all the ranges:
0.003% of reading + 2 counts
Accuracy:
0.003% of reading + 0.0023 Hz
0.003% of reading + 0.0336 Hz
0.003% of reading + 0.138 cpm
0.003% of reading + 0.5 cph
0.0011% FS / °F
(0.002% FS / °C)
Unipolar
Bipolar
0 ... 30 V
0 ... 24 V, Resolution: 0.1 V
0 ... 24 V dc ± 1%
0 ... 24 V ac ± 5% (Current ≤ 20 mA)
Four 0.16 in (4 mm) sockets
0 to ±55 mA
0.02% of reading + 3 counts
0.0011% FS / °F
(0.002% FS / °C)
Open and closed. 2 mA current.
24 V ± 10%
250 Ω (menu selection)
Three 0.16 in (4 mm) sockets
K395 Issue 3