NOTICE
• To reduce risk of damage and ensure an accurate result
with the 3M™ Clean‑Trace™ Luminometer LX25:
− Do not operate the Clean-Trace Luminometer LX25 in
close proximity to strong magnetic fields .
− Do not store or operate the Clean-Trace
Luminometer LX25 at temperatures above 70°C .
− Do not store or operate the Clean-Trace
Luminometer LX25 at temperatures below -40°C as
condensation may occur . The instrument should be
allowed to warm to room temperature before use
and a cloth or tissue should be used to wipe dry the
outer surfaces .
− Only insert 3M™ Clean-Trace™ ATP Tests into the
Clean-Trace Luminometer LX25 chamber .
− Do not use cleaning agents in the Clean-Trace
Luminometer LX25 chamber .
− Ensure the ATP Test is inserted into the Test sleeve and
activated before inserting the Test into the Clean-Trace
Luminometer LX25 chamber .
− Do not tilt the Clean-Trace Luminometer LX25 while
ATP Tests are in the chamber .
Environmental Operating Conditions
Environmental Condition
Indoor Use only
Altitude
Operating Temperature Range
Relative Humidity
Storage Temperature Range
AC voltage for power supply
Mains supply voltage fluctuations up to +10% of
nominal voltage
Temporary overvoltages occurring on the mains supply
Frequency
AC Current for power supply
Power supply output voltage
Power supply output current
Input current
Overvoltage
Pollution Degree
− Do not leave a used (activated) ATP Test in the
Clean-Trace Luminometer LX25 chamber after the
result has been read and recorded .
− Ensure annual recommended service and calibration of
the Clean-Trace Luminometer LX25 is completed by the
3M Service Center or 3M authorized service personnel .
− Periodically inspect the Clean-Trace
Luminometer LX25 for damage, including cracks
or swelling .
− Do not store or use the Clean-Trace
Luminometer LX25 or power supply in a location
exposed to steam, high humidity, dust, or
strong vibration .
− Reference the Environmental Operating Conditions
table in the Important Safety Information for
Clean-Trace Luminometer LX25 in this User Manual .
− Use only 3M Clean-Trace ATP Tests and 3M accessories
supplied by your local 3M sales representative
or distributor .
− Do not use the Clean-Trace Luminometer LX25 if liquid
or a foreign object gets inside the device .
Operating Condition
- Only use the data connections to a
computer that conform to IEC/EN/UL/
CSA 60950-1
- Keep Clean-Trace Luminometer LX25
away from liquids and high humidity .
3000 maximum
+5 to +40 (+41 to +104)
10 – 90 Non-condensing
-40 to +70 (-40 to +158)
100-240
50-60
300
5
1 .2
850
Category II
2
4
Units
Not applicable
Meters
°C (°F)
%
°C (°F)
Volts
Hertz
mA
Volts DC
Amps
mAMPs
Not applicable
Not applicable