4.
Motor and Electrical Maintenance: The Model 614M motor should not
need servicing for the life of the centrifuge. The electrical components are
selected for high reliability and should not need service.
5.
Tube Holder Replacement: It is recommended that the tube holders be
replaced after 24 months of use. Inspect tube holders regularly for cracks. If
cracks are discovered, replace immediately.

Cleaning and Disinfection:

To prolong the life of the centrifuge cleaning and disinfection is recommended every
six months, or whenever there is a spillage or tube breakage. Contaminants must be
removed immediately, or corrosion and premature degradation of components can
occur.
a)
Unplug the centrifuge before cleaning.
b)
Autoclaving of any component of the centrifuge is prohibited.
c)
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge
in water or other cleaning solutions as this will cause damage and void the
warranty.
d)
ONLY isopropyl alcohol, soap and water, or a 10% (5500 PPM) bleach solution
should be used for cleaning and disinfection of the centrifuge and accessories.
e)
All surfaces must be dried immediately after cleaning and disinfecting.
f)
TBQ Germicidal products shall not be used, as they will cause damage to the
centrifuge and void the warranty.
g)
The use of fully/partially halogenated hydrocarbons, ketones, esters, ethers,
benzyls, ethyl benzenes, and all other chemicals not prescribed by the
manufacturer shall not be used as they may cause damage to the rotor
chamber, rotor, tube holders, accessories and centrifuge exterior and void the
warranty.
h)
It may be necessary to remove the rotor and clean the rotor chamber. Refer to
the Installation and removal of the rotor section of this manual.