6.0 Verification Procedure

This procedure verifies the positive/negative voltage detection levels and ground
monitoring performance of the 3M
Specifications of this user's guide.
Equipment Required – Supplied by User
• DC Power Supply (DCPS) with connecting wires. Output 0 to +10VDC
minimum.
• DC Voltmeter (DCV) with digital display ± 1% accuracy and test leads. Use of
a voltmeter is only required if DCPS does not include a digital display.
• Resistance Substitution Box (RSB), 1Ω to 12Ω (minimum) ±1% or discrete
resistors 8 and 12Ω ±1% value.
3M
Verification Kit 790VK (Ordered Separately)
• 3M
Dual Conductor Ground Cord 2360 (5 ft.).
• Two-wire male connector with 24" tinned lead-wires.
• Test wrist band socket.
Procedure:
1. Ground Disconnect / Hi & Lo Audible Alarm Volume Level Functions
Ground Disconnect Specification: 10Ω ±20%
a. Connect the RSB to the two-wire male connector with 24" tinned lead-
wires. Insert the two-wire male connector into the mating jack located at
the rear of the monitor.
b. Connect the AC adaptor supplied with the Static Monitor 790 to the
25VDC input power-jack located at the rear of the monitor and plug the
AC adaptor into a wall outlet.
c. Set the RSB as follows and observe all lamps and audible alarm:
8Ω – Green lamp ON, Red lamps OFF, and audible alarm OFF.
12Ω – Green lamp OFF, Red lamps both ON (Continuous illumination),
and audible alarm ON (Continuous tone).
Note: Confirm Hi and Lo audible alarm volume levels during 12Ω test condition by adjusting
Hi & Lo slide switch located at the rear of the monitor while alarm is sounding.
2. Wrist Strap Function – Open Circuit Test
a. Adjust the RSB connected to the two-wire male connector to 8Ω or less
to silence the Ground Disconnect alarm. The green lamp should be ON at
this time.
Static Monitor 790 as stated in Section 7,
12