When the AUTO OFF feature is enabled, the stethoscope will enter standby mode after a period of time determined by the
POWER setting if there is no contact detected on the diaphragm AND the buttons are not activated. When the AUTO OFF
feature is disabled, the stethoscope will enter standby mode after a period of time determined by the POWER setting from
the last button activation. In addition, when the AUTO OFF feature is disabled, there will be an audible alert 10 seconds prior
to entering standby mode. After entering standby mode, the Model 3100 will be powered on when contact is detected on the
diaphragm or by button activation. Otherwise, the Model 3100 will completely power off after the period of time determined
by the POWER setting (see below).
With AUTO OFF feature ENABLED:
SETUP
Options
POWER
HIGH
MEDIUM
LOW
With AUTO OFF feature DISABLED:
SETUP
Options
POWER
HIGH
MEDIUM
LOW
Detail
Remains on full power for 30 seconds before entering standby
Remains in standby for 5 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
Remains on full power for 20 seconds before entering standby
Remains in standby for 2 hours before powering off
Backlight remains lit for 5 seconds after pressing the power button
Remains on full power for 10 seconds before entering standby
Remains in standby for 30 minutes before powering off
Backlight remains lit for 3 seconds after pressing power button
Detail
Remains on full power for 7 minutes before entering standby
Remains in standby for 5 hours before powering off
Backlight remains lit for 5 seconds after pressing power button
Remains on full power for 5 minutes before entering standby
Remains in standby for 2 hours before powering off
Backlight remains lit 5 seconds after pressing power button
Remains on full power for 3 minutes before entering standby
Remains in standby for 30 minutes before powering off
Backlight remains lit 3 seconds after pressing power button
12