DSC ELECTRONICS DP-D Series 사용자 설명서 - 페이지 9

{카테고리_이름} DSC ELECTRONICS DP-D Series에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. DSC ELECTRONICS DP-D Series 14 페이지. High power laboratory power supplies

DSC ELECTRONICS DP-D Series 사용자 설명서
DP-D Series User's Manual
Version 08.01.2019
Operation via the front panel:
1). Switch to PRESET mode by pressing the PRESET button.
2). Press the CURRENT rotary encoder, the OV LED lights up.
3). Now set the maximum voltage by turning the VOLTAGE rotary encoder.
4). Confirm the entry by pressing the CURRENT rotary encoder.
Setting via digital interfaces:
The command 0x2d 0x02 directly addresses the voltage limitation. The command is an extension of the DP-D Series
communication protocol and is structured as follows (XX YY YY ZZ ZZ AA AA) where X is the device ID, Y is the
command, Z is the value and A is the CRC sum. The command to set the voltage limit to 5.11V would thus be for a
device with the ID 01:
01 2D 02 01 FF F0 BC
ID: 01 / BEFEHL: 2D 02 / WERT: 01 FF / CRC: F0 BC
DSC-Electronics Germany • Georgstraße 36 • 53111 Bonn
DSC-Electronics Germany • Georgstraße 36 • 53111 Bonn