Frequency to DC Transmitters, Isolated, Factory Configured
Input:
0-25 Hz to 0-20 kHz
Output:
0-1 V to 0-10 VDC, ±1 VDC to ±10 VDC, 0-1 mA to 20 mADC
O Precision Frequency to DC Converter
O Zero and Span Output Calibration Potentiometers
O Full 1200 V Input/Output/Power Isolation
O Input and Output LoopTracker
LEDs
®
O Output Test Button
O Built-In Loop Power Supply for Sink/Source Output
Applications
Q Monitor and Control Motor or Line Speed
Q Convert Speed and Frequency Signals
Q Simplify Interfacing of Frequency Sensors
Frequency Input Range
Factory configured, please specify input range
Frequency: 0-25 Hz to 0-20 kHz
Sine wave, sawtooth, or square wave with min. 5 µsec pulse
Input Amplitude Range
100 mV to 150 V
RMS
Any waveform with minimum 100 mV amplitude change
Input Impedance
10 kW nominal (maximum sensitivity)
100 kW nominal (minimum sensitivity)
Input Sensitivity/Hysteresis
Multi-turn potentiometer for sensitivity adjustment
Maximum sensitivity:
±25 mV typical
Minimum sensitivity:
±2.5 V typical
Input Protection
Normal mode:
200% of input rating
Common mode:
600 VDC or 600 VAC
Input Power Supply
15 VDC ±10%, regulated, 25 mADC
Max. ripple, less than 10 mV
RMS
May be used to power sensor
LoopTracker
Variable brightness LEDs indicate I/O loop level and status
DC Output Range
Factory configured, please specify output range
Voltage:
0-1 VDC
to
Voltage, M09 option:
0-10 VDC
to
Bipolar voltage:
±1 VDC
to
Current:
0-2 mADC
to
20 V compliance, 1000 W at 20 mA
Output Calibration
Multi-turn zero and span potentiometers, ±15% adj. range typ.
Output Loop Power Supply
20 VDC nominal, regulated, 25 mADC
Max. ripple, less than 10 mV
RMS
May be selectively wired for sinking or sourcing mA output
Output Test/Override
Front momentary button or external contact closure sets out-
put to test level. Potentiometer adjustable 0-100% of span.
Output Ripple and Noise
Less than 10 mV
ripple and noise
RMS
Linearity
Better than ±0.1% of span
Ambient Temperature Range and Stability
–10°C to +60°C operating ambient
Better than ±0.02% of span per °C stability
Isolation
1200 V
minimum
RMS
Full isolation: power to input, power to output, input to output
Housing and Connectors
IP 40, requires installation in panel or enclosure
For use in Pollution Degree 2 Environment
Mount vertically to a 35 mm DIN rail
Four 4-terminal removable connectors, 14 AWG max wire size
Power
85-265 VAC, 50/60 Hz or 60-300 VDC, 2.5 W maximum
D versions: 9-30 VDC or 10-32 VAC 50/60 Hz, 2.5 W max.
A
A
P
P
BSOLUTE
BSOLUTE
ROCESS
ROCESS
File E145968
85-265 VAC, 60-300 VDC
model only
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
H H H H H
H H H H H H
Made in USA
Applications Link
api-usa.com/apps
Free Factory
I/O Setup!
input to ground
p
Dimensions
0.89" W x 4.62" H x 4.81" D
22.5 mm W x 117 mm H x 122 mm D
Height includes connectors
0-10 VDC
Description
0-20 VDC
The APD 7010 accepts a frequency input and provides an
±10 VDC
optically isolated DC voltage or current analog output that is
0-20 mADC
linearly proportional to the input.
The full 3-way (input, output, power) isolation between
input and output makes this module useful for ground loop
elimination, common mode signal rejection, and noise pickup
reduction.
Also standard on the APD 7010 is a 15 VDC input excitation
supply. If needed, this supply can be used to power a passive
sensor, eliminating the need for an additional external power
supply.
Common applications include frequency to DC conversions
from frequency output type devices such as rotary encoders,
magnetic pick-ups, variable speed drives and flow meters.
A PLC pulse rate output can be programmed to vary. By con-
necting the APD 7010 to this output, a proportional analog
signal can be generated.
How to Order
All models are factory configured.
Order APD 7010 D for operation on low voltage power
Milliamp outputs can be field wired for sink or source.
Model
APD 7010
APD 7010 D
Options–add to end of model number
M01
Input/output reversal, such as 20-4 mA output
M09
High voltage output up to 20 V (specify range)
U
Conformal coating for moisture resistance
I
I
NSTRUMENTS
NSTRUMENTS
Connect mA Output
for Sink or Source
Adjustable Output
5 6 7 8
Test Function
Adjustable Input
Sensitivity
Input LoopTracker
LED
Zero and Span for
Output
Output LoopTracker
LED
15 VDC Sensor
Power Available
9 10 11 12
Lead
Pb
Free
Universal
Power
Input
Factory ranged
0-25 Hz to 0-20 kHz
1220 American Way Libertyville, IL 60048
Phone:
800-942-0315
APD 7010
Quick Link:
1 2 3 4
13 14 15 16
Sink/Source Versatility
For maximum versatility the output can be selectively wired for
sinking (unpowered) or sourcing (powered) milliamp output.
The 20 VDC loop excitation supply can be used to power a
milliamp current loop if required. The output can also be wired
for an externally powered loop.
LoopTracker
API exclusive features include two LoopTracker LEDs (green for
input, red for output) that vary in intensity with changes in the
process input and output signals. These provide a quick visual
picture of your process loop at all times and can greatly aid in
saving time during initial startup and/or troubleshooting.
Output Test
An API exclusive feature includes the test button to provide a
fixed output (independent of the input) when held depressed.
The test output level is potentiometer adjustable 100% of the
output span.
The output test button greatly aids in saving time during initial
startup and/or troubleshooting.
Please specify
Model
Input range in Hz
Output range
Options as required
Output
85-265 VAC or 60-300 VDC
Factory ranged
voltage or milliamps
9-30 VDC or 10-32 VAC
Accessory—order as separate line item
API BP4 Spare removable 4 terminal plug, black
api-usa.com
Fax: 800-949-7502
api-usa.com/7010
Removable Plugs
See Wiring
Diagrams on
Next Page
Power
© 02-20