DTE Lux E 사용 설명서 - 페이지 9
{카테고리_이름} DTE Lux E에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. DTE Lux E 16 페이지. Curing light
2.0bar~2.3bar (0.20MPa~0.23MPa) before each use, other parts should
be cleaned by clean water or neutral sterilized liquid, but do not soak
the equipment in the water. Do not clean by volatile or soluble liquid,
otherwise the marks of the control panel will fade.
10.2 Please clean the optical fiber to avoid the remaining resin on the
surface and infect the life-span and the effectiveness of solidification.
11. Troubleshooting
Faults
No indication
No response.
The screen shows
"Er".
The screen shows
"E1".
Light intensity is
weak.
The equipment
d o e s n 't c h a rg e
when the adapter
is connected.
Possible causes
1. Battery is out of
power.
2. Faulty of battery.
3. Battery is protected.
Faulty of main unit.
Low battery.
1. The optical fiber is
not installed correctly.
2. There is crevice on
the optical fiber.
3. There is resin on
the top of the optical
fiber.
1. The adapter is not
connected well
2. Faulty of adapter or
incompatible.
3. The charging point
is impurity.
7
Solutions
1 . C h a n g e a n e w
battery.
2 . C h a n g e a n e w
battery.
3. Charge.
Repair.
Charge.
1 . R e i n s t a l l t h e
optical fiber.
2 . C h a n g e a n e w
optical fiber.
3. Clean the resin.
1. Reconnect.
2. Change the adapter.
3 . C l e a n b y t h e
alcohol.