Do not immerse the probe casing.
Do not clean the probe with acetone, ethyl alcohol or sodium hypochlorite (bleach) as this will
damage the surface
Avoid any cleaner that may scratch or dissolve plastic surface
High level disinfection using liquid agent.
The probes can be sterilised with peracetic acid solution (for example Anioxyde).
Soaking duration: 20 min
Leak tight length of the probes: 20 mm
After cleaning
The proper state of the probe must be checked after sterilisation: a visual checking of the extremity of the
probe must be performed. There must not be any cracks or upheaval.
7.4

BATTERY REPLACEMENT

8

Service

8.1

Microflow S spare parts

Order number designation
04 739
Loud speaker
04 753
Microflow grid
08 470
Casing
08 512
Electronic board
8.2

Probe warranty

The pencil probes are warranted for six (6) months against manufacturing defect.
Limitation of the duration of the probe warranty in the case of liquid sterilisation
The warranty of the pencil probes is limited to 50 sterilisation periods of 20 min each.
8.3

Assembly

20 174 C
Microflow-S pocket Doppler
User's Documentation
The battery should be replaced as soon as the yellow
LED flashes
Respect the battery type: 9 V alkaline - 6LR61 or PP3
Be careful about the polarity when replacing the
battery.
Order number designation
11 830
11 837
04 746
Microflow S: keyboard
Labels
Plastic box
Page 10/12