Anyload A2P 제품 설명서 - 페이지 5
{카테고리_이름} Anyload A2P에 대한 제품 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Anyload A2P 6 페이지. Load cell amplifier
5.
Output Mode Switch
6.
Calibration
The calibration of A2P consists of Zero Calibration and Span Calibration:
6.1 Zero Calibration
Step1.Remove all load from the scale platform. If the scale require hooks or
chains (tare weight), place the hooks or chains onto the scale for zero
calibration.
Step2.Adjust ZERO variable resistor to an output of 0V or 4mA.
(Note: Tare weight shall not exceed 30% of full load)
6.2 Span Calibration
Step1.Place full load onto the scale.
Step2. Adjust SPAN variable resistor to an output of 10V or 20mA.
(Note: It's recommended to repeat adjustment in Step 2 of Section 6.2 above
three times.)
4 |
ANYLOAD A2P Amplifier Product Manual (V1611)