Anyload LCT100 사용자 설명서 - 페이지 3
{카테고리_이름} Anyload LCT100에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Anyload LCT100 4 페이지. Load cell
If full testing cycle was chosen:
The LCD shows " Test in progress" and the LED blinks for a few seconds.
Once the test is finished you may scroll the results by pressing the up arrow
key. The results will be displayed in the following order:
Input – Resistance between Ex+ and Ex- ( Max 5000Ohm )
Output - Resistance between Signal+ and Signal- ( Max 5000Ohm )
Sense+ - Resistance between Ex+ and Sense+ ( applicable for 6 wires load
cell ) ( Max 5000Ohm )
Sense- - Resistance between Ex- and Sense- ( applicable for 6 wires load cell
) ( Max 5000Ohm )
Signal Output – Load cell signal in percentage of full L/C capacity. Note: in
case of a broken cable or a very high input/output resistance the display
will show " Cannot measure"
Shield to Brdg – Resistance between the cable shielding ( screen ) and the
bridge
Body to Brdg – Resistance between the load cell body and the bridge
Shield to Body – Resistance between the load cell body and the cable's
shielding
Tips for trouble shooting Load Cell and scales
Load cells might be damaged due to many reasons: Overloading, lightning
strikes, moisture or chemical ingress, vibrations, corrosion, pulling the cable etc.
As a result, the scale may provide unstable readings, zero drifting, error reading,
non-linearity etc.
The LCT100 is the only measuring tool required to perform the tests either on
site or in the lab. Before you start, make sure to have the Load cell spec and the
cable color code (usually you get it with the load cell, or it can be downloaded
from the manufacture web site).
How to test a weighing platform with a few load cells:
1. First stage: Leave all load cells connected to the summing box.
Disconnect the cable from the weighing indicator, connect it to the
LCT100 and perform a test.
2. Expected results for a good scale are as per the following example: If
you have 4 load cells of 1000Ω, 100kg each and 40kg are applied on
the scale than you would expect to see around 250Ω for the input and
2 |
ANYLOAD LCT100 Load Cell Tester User's Guide (V1702)