System S1P (pod) only
WARNING
• Always place the scanner in its intended mount on the
pod when not in use.
• Do not place the pod on a slope greater than 5 degrees.
• Place all cables carefully so that patient and/or user
do not get caught in them and accidentally pull any part
of the system. Pulling cables may damage the system,
interrupt the data connection, and/or drag the system
off its supporting surface.
5.3 Explosion Hazards
Environment
WARNING
The TRIOS IOS system is not designed to be used
in potentially explosive environments, such as in close
proximity to flammable liquids or gases, or in oxygen-
enriched atmospheres.
Distance to the patient with anesthetic
WARNING
There is a potential explosion hazard if the TRIOS IOS system
is operated in the presence of flammable anesthetic.
5.4 Electrical Safety
Ensure a grounded/earthed connection
WARNING
To avoid the risk of electric shock, the MOVE/MOVE+ cart
must only be connected to a supply mains with grounding/
protective earth. The MOVE/MOVE+ cart is equipped
with a three-pin (US: prongs) grounding/earth type plug.
If you cannot insert the plug into your outlet, contact
your electrician to replace the plug or outlet, to ensure
a grounded power connection. Do not try to defeat or
ignore the purpose of the grounding-type plug.
Electrical shock
WARNING
There is a risk of electrical shock if you attempt to access
the inside of any part of the system. Only authorized and
qualified service personnel may access the inside of any
part of the system.
Do not use a power strip
WARNING
Do not connect the TRIOS IOS system to an ordinary power
strip or extension cord, because these may not be secure in
relation to grounding. If the ground connection fails, these
can cause the following hazards and potential injury:
• The total leakage current for all the connected equipment
can exceed the limits specified in EN (IEC) 60601-1.
• The impedance of the ground connection could exceed
the limits specified in EN (IEC) 60601-1.
8
Stress on cables
CAUTION
All externally connected cables (the power cable,
network cable, and the handheld scanner cable)
must never be subjected to pulling stress.
Spilled liquids
WARNING
Do not bring liquids, such as beverages, near the TRIOS
IOS system. Do not spill liquids on the TRIOS IOS system.
Condensation
WARNING
Variations in temperature or humidity may cause water
to condense inside the system, which can cause damage.
Always let the system reach room temperature before you
plug it into the power source.
If the system has been subject to major temperature or
humidity changes, wait till the system has reached room
temperature before plugging it into the power source.
If there is visible evidence of condensation, wait at least
8 hours before plugging it into the power source, to ensure
that no internal condensation occurs.
Disconnecting from mains
WARNING
There is no power ON/OFF switch on the system;
therefore, the only reliable means to disconnect the system
from mains is to unplug the power cord. Do not position
the system so that it is difficult to unplug the power cord.
Please allow enough time for the system to shut down
properly before disconnecting the power cord.
System S1P (pod) only: You must also disconnect the PC
from mains.
PC connection options and requirements
System S1P (pod) only
WARNING
The PC is a standard IT equipment certified to IEC 60950-1
and is located at least 1.5 m away from the patient. If you
choose this setup, you must not scan the patientand touch
the PC at the same time! You must not touch any peripherals
connected with a cable to the PC while scanning a patient.
See the PC's instructions for use for further details.
WARNING
Only use the power supply that is delivered as part of the
system. OTHER POWER SUPPLIES MAY BE UNSAFE AND/OR
MAY DAMAGE THE TRIOS IOS SYSTEM!