Specification
Type:
• CEL-120/1: Calibrator to EN (IEC) 60942: 2003 Class 1
• CEL-120/2: Calibrator to EN (IEC) 60942: 2003 Class 2
Both calibrators meet ANSI S1.40 - 2006
• Calibration reference condition: 23 °C, 101.3 kPa and 50 %RH.
Reference Microphone:
• B & K 4133, Nominal load volume: 190 mm²
Calibration level: (at reference conditions)
• CEL-120/1: 114.0 dB ±0.25 dB and 94.0 dB ±0.25 dB,
• CEL-120/2: 114.0 dB ±0.35 dB.
• Calibration frequency: 1 kHz ±5 Hz.
• Stabilising time:
5 s
• Sensitivity to load volume change: 0.0002 dB/mm²
Short term level stability:
• < ± 0.1 dB for CEL-120/1
• < ± 0.2 dB for CEL-120/2
• Total harmonic distortion: Less than 1%.
Tolerance limits for the CEL-120/1 (Class 1) over full environmental conditions:
• Static pressure range: 650 to 1080 mBar
• Air temperature range: –10 to +50 °C
• Relative humidity range: 25 to 90 % RH
• Pressure coefficient: <0.004 dB/kPa
• Better than <±0.4 dB deviation over environmental ranges
Tolerance limits for the CEL-120/2 (Class 2) over full environmental conditions:
• Static pressure range: 650 to 1080 mBar
• Air temperature range: –10 to +50 °C
• Relative humidity range: 25 to 90 % RH
• Pressure coefficient: <0.004 dB/kPa
•
Better than <±0.6 dB deviation over the environmental range
Storage temperature range: –20 to +60 °C.
Battery: 2 × 1.5 V, AAA (LR03).
An alkaline battery is recommended and typically provides 2 minutes of operation per
day for almost 2 years.
Dimensions: 75.5 mm × 54.0 mm × 37.0 mm (2.9 × 2.1 × 1.4 in).
Weight (including batteries): 85 g (0.19 lb).
Manufacturers Servicing and Warranty Arrangements
In order to ensure its rigid conformity with the requirements of the specification, this
instrument is thoroughly inspected and calibrated prior to dispatch from the factory.
All technical information for an individual instrument is filed under the instrument
serial number. Therefore, the instrument serial number should be quoted in any
correspondence concerning the instrument.
The manufacturers undertake to rectify any defect in the instrument that is directly
attributable to faulty design or assembly, and which becomes apparent during the
warranty period. In order to take advantage of this warranty, the instrument must be
returned, carriage paid, to the manufacturer's factory or accredited agent, where
necessary repairs will be carried out.
Normally, the warranty period runs for 24 months from the date of receipt of goods,
with exceptions on certain specialised components supplied by other manufacturers
which are warranted for shorter periods. Some of the specialised components used in
this instrument may be subject to longer guarantees by their actual manufacturers
and in all such cases the benefits of these undertakings will be passed on to the user.
However, Casella CEL's liability is limited to items of their own manufacture, and they
do not accept liability for any loss resulting from the operation or interpretation of the
results from this equipment.
A comprehensive Instrument Calibration Maintenance Agreement (ICMA) scheme is
available to extend the initial warranty period of this instrument. At the end of the first
warranty period, it is recommended that the equipment be returned to the Service and
Re-calibration Department at Bedford, where it will be inspected and entered into the
ICMA scheme as required. The warranty will then be extended for the period stated
on the individual schedule. Please contact your local Casella CEL agent for full details
of this service.
In the event of a malfunction developing during the warranty period, the instrument
should be carefully packed and returned either to Casella CEL's local agent, or in the
case of domestic sales, to the Service Department at Bedford.
Please include the following information:
• Instrument type(s) and serial number(s)
• Customer name and address
• Contact name and phone number
• Reason for returning the equipment, with a detailed description of the fault
The necessary adjustments or repairs will be carried out, and the instrument returned
as soon as possible. After the warranty has expired (except on approved accounts)
service work is undertaken against quotations, and all packing and transit costs are
charged extra.
Introduction
The CEL-120/1 and CEL-120/2 Acoustic Calibrators provide an accurate and easily
used means of calibrating sound level meters, both in the laboratory and in the field.
CASELLA MEASUREMENT
Regent House, Wolseley Road, Kempston, Bedford, MK42 7JY, U.K.
Phone: +44 (0) 1234 844 100
In addition, both instruments meet the requirements of ANSI S1.40 - 2006.
Fax: +44 (0) 1234 841 490
E-mail for technical support: [email protected]
These calibrators are intended for use with industry standard WS2 (half-inch)
For online Technical Support: http://helpdesk.casellameasurement.com
microphones, and may also be used to calibrate WS3 (quarter-inch) microphones by
means of the CEL-4726 Coupler.
© 2011 Casella Measurement
Document number FG43-02
Field Guide
CEL-120/1 and CEL-120/2
Acoustic Calibrators
4
1
2
5
3
6
1. On/Off button
4. Upper blue LED (114 dB level)
2. Green LED power status
5. Level toggle button
3. Microphone cavity
6. Lower blue LED (94 dB level)
• The CEL-120/1 satisfies the requirements of EN (IEC) 60942:2003 Class 1 at
user-selected calibration levels of 94.0 dB and 114.0 dB.
• The CEL-120/2 satisfies the Class 2 standard at 114.0 dB.