Freedom iQ Vacuum App For Prosthetists (iQ Vac App)

Dashboard

After connecting to an iQ Vacuum unit, an information Dashboard
summarizes key setup and performance indicators. The live graph
at the bottom of the screen shows you how the socket vacuum is
changing and how the vacuum is responding to the live changing
activity of the user.

Adjust Vacuum Levels

While iQ Vacuum is fully automatic, the amount of vacuum targeted
at different activity levels is adjustable by the prosthetist by selecting
Controls>Settings>Vacuum Level.

Initial Training

For optimal automatic function, iQ Vacuum must be trained to
recognize the unique gait of the user. Before training, you may choose
to alter the standard target vacuum levels. In the Freedom iQ Vacuum
select Controls>Settings>Step Detection Setup. The app will lead you
through the process of tuning the iQ Vacuum sensor.
The process can be repeated by the prosthetist at any time in the
future should there be a question about the performance of the
adaptive control system, or if the target vacuum levels are changed.
12
Freedom Innovations |
iQ Vacuum
| Instructions for Use

Step Detection Setup

To use Step Detection Setup, have the user walk continuously. Each
time that the device observes a step, the gray circle in the middle will
flash green and there will be an audible beep. Make sure the sound
on your mobile device is turned on and the volume is up. There should
be one flash/beep for each step you observe. Ignore starting, stopping
and turning steps. After watching the user for several steps, tap the
control on the screen that reflects if the device detected:"Too Few",
"Too Many" or "Just Right" number of steps. Continue the process until
it is "Just Right". Tap "Just Right" at the bottom of the screen. Be sure
to observe the user at a variety of cadences.

Adjust Target Step Rates

The vacuum level target increases with increasing activity levels
(Low, Medium, High) as defined by the target step rates. These are
adjustable by selecting Controls>Settings>Step Rate in the app menu.
You may want to increase the rate if you feel the device is going into
medium or high activity mode too early or you may want to decrease
the rates if the user does not achieve the higher activity levels.
13