4.3 Product cleaning & care guidelines
The X350 features BioProtect® antimicrobial coating on all painted surfaces and has a fully-
enclosed head with an Ingress Protection (IP) rating of 54.
These guidelines are provided to help you maintain your equipment properly and prevent any
damage. Use of incompatible cleaning chemicals can adversely affect the finish and physically
damage the metal and plastic parts of the DARAY product, invalidating its warranty.
Please ensure these instructions are passed to the persons responsible for cleaning and they are
followed before each use.
Manufacturers recommended cleaning method:
1.
Turn off the power to the light, and where possible, unplug the mains socket.
2.
Remove noticeable contamination. Pay special attention to the light head and handle.
3.
Ensure the equipment is surface-cleaned and disinfected using an appropriate, non-
abrasive cleaning agent (see recommended list below).
4.
Dry all surfaces thoroughly with a clean cloth and check for damaged surfaces,
discolouration and/or contamination.
NOTE: Some DARAY products feature our BioProtect® antimicrobial coating on painted finishes.
This is designed to assist with infection control and is not a replacement for cleaning.
CAUTION: Our products are NOT waterproof. The equipment and accessories (included chargers
and battery packs) MUST NOT at any point be saturated, to prevent water ingress. Use a damp
cloth or saturated wipe if necessary. All electrical safety guidelines should be followed.
Any damage caused as a result of not following these guidelines will void any product warranty.
Below is a partial list of recommended surface agents:
Cavicide™ products
Isopropyl alcohol-based cleaners (<40% alcohol)
Ethyl alcohol-based cleaners (<40% alcohol)
Lysol® disinfectant
Other non-bleach and non-abrasive disinfectants or cleaners
DO NOT USE agents containing the following:
Organic, mineral and oxidizing acids (min. pH value 8.5)
Bases (max. acceptable pH value 8.5)
Oxidation agents (e.g. Hydrogen Peroxide)
Halogens (chlorine, iodine, bromide)
Aromatic/halogenized hydrocarbons
Bleach-based cleaners (e.g. Clorox™, Sterilox™)
Abrasive cleaners (e.g. Comet Cleaner™)
Acetone-based cleaners (e.g. nail polish remover, Goo-off™)
MEK (Methyl Ethyl Ketone)
If you are unsure whether you are using the correct cleaning fluids on your DARAY equipment,
please feel free to get in touch with us for advice.
Page 12 of 17