6.
INSTALLATION
6.1. General recommendation
The depth probes are installed in places where the liquid level is measured in wells, tanks, boreholes,
etc. The probe is immersed in the measured medium. A special cable extends above the medium
level and can be connected directly to the device cooperating with the probe or to the junction box.
The probe can be hung on the power cable, e.g. by using Aplisens SG cable hanger (item 1
in Fig. 1). In the event of frequent removal of the probe or when there is a risk of catching on
protruding elements during pulling up, it is recommended to hang the probe on a steel cord using
the carrying eye (item 2 in Fig. 1 - does not apply to SGE-16 probes).
Figure 1. Mounting brackets for probes.
The probe with an additional Teflon shielding on the cable should be hung on a carrying cable or
on an internal cable (do not grab the Teflon shield). Hang the probe in Ex version with the
grounding cord only by the lifting eye on an additional supporting wire.
Immediately before placing the SGE-25S or SGE-25C probe in the measured medium,
remove the protective cover from the probe separator. During installation, protect the
probe against mechanical shocks.
Figure 2. Probes diaphragm cover for SGE-25C and SGE-25S.
A special cable extends above the medium level and can be connected directly to the device
cooperating with the probe or to the junction box. If the probe is to be in the current or in the
turbulence area, it should be installed in a protective pipe, e.g. made of PVC.
Revision 01.A.001/2021.09
EN.IO.SGE
9