Integer Registers
54
55
56
57
58
59
60
61
62
63
64
65
*Registers marked with an asterisk (*) cannot be written to unless a specific procedure is followed. Contact
Apogee Instruments to receive the procedure for writing these registers.
Write only registers (function code 0x10).
Integer Registers
Packet Framing:
Apogee sensors use Modbus RTU packets and tend to adhere to the following pattern:
Slave Address (1 byte), Function Code (1 byte), Starting Address (2 bytes), Number of Registers (2 bytes), Data
Length (1 byte, optional) Data (n bytes, optional)
Modbus RTU packets use the zero-based address when addressing registers.
For information on Modbus RTU framing, see the official documentation at
http://www.modbus.org/docs/Modbus_Application_Protocol_V1_1b3.pdf
baudrate (0 = 115200, 1 = 57600, 2 = 38400, 3 = 19200, 4 = 9600, any other
190
PAR offset*
Far-red multiplier*
Far-red offset*
running average
heater status
slave address
model number*
serial number*
number = 19200)
parity (0 = none, 1 = odd, 2 = even)
number of stopbits
PAR multiplier (shifted two decimal points to the left)*
PAR mV offset (shifted four decimal points to the left)*
Far-red multiplier (shifted two decimal points to the left)*
Far-red mV offset (shifted four decimal points to the left)*
running average
heater status
Writing to this register resets Coefficients to firmware
defaults. (NOT factory calibrated values!) Slave Address =
1, Model = 442, Serial = 1000, Baud = 3, Parity = 2, Stopbits
= 1, running average = 1