Apogee Instruments SU-202-SS 소유자 매뉴얼 - 페이지 14

{카테고리_이름} Apogee Instruments SU-202-SS에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Apogee Instruments SU-202-SS 18 페이지. Ultraviolet sensor
Apogee Instruments SU-202-SS에 대해서도 마찬가지입니다: 소유자 매뉴얼 (16 페이지)

Apogee Instruments SU-202-SS 소유자 매뉴얼

MAINTENANCE AND RECALIBRATION

Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated
materials on the diffuser of the SU-200 series UV-A sensors can block the optical path in three common ways:
1. Moisture or debris build-up on the diffuser.
2. Dust during periods of low rainfall.
3. Salt deposit accumulation from evaporation of sea spray or sprinkler irrigation water.
SU-200 series sensors have a domed diffuser and housing for improved self-cleaning from rainfall but active
cleaning may be necessary. Dust or organic deposits are best removed using water, or window cleaner, and a soft
cloth or cotton swab. Salt deposits should be dissolved with vinegar and removed with a cloth or cotton swab.
Never use solvents such as alcohol or acetone to clean the sensor. Use only gentle pressure when cleaning the
diffuser with a cotton swab or soft cloth, to avoid scratching the outer surface. The vinegar should be allowed to
do the cleaning, not mechanical force. Never use an abrasive material or cleaner on the diffuser.
It is recommended that UV-A sensors be recalibrated every two years. See the Apogee webpage for details
regarding return of sensors for recalibration
repairs/).
(http://www.apogeeinstruments.com/tech-support-recalibration-