Apogee Instruments SU-300 UV-B 소유자 매뉴얼 - 페이지 8
{카테고리_이름} Apogee Instruments SU-300 UV-B에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Apogee Instruments SU-300 UV-B 19 페이지. Ultraviolet index & ultraviolet-b sensor
Cosine Response
Cosine Response Graph
10
8
6
4
2
0
-2
-4
-6
-8
-10
0
10
20
30
40
50
60
Solar Zenith Angle [°]
Directional, or cosine, response is
defined as the measurement error at
a specific angle of radiation incidence.
Directional error for Apogee SU-300
series UV-I/UV-B sensors is
approximately ± 2 % and ± 5 % at solar
zenith angles of 45° and 75°,
respectively.
Mean cosine response of four
Apogee UV-I/UV-B sensors. Cosine
response was calculated as the
relative difference of UV-I/UV-B
sensors from a reference UV-I/UV-B
sensor deployed outdoors. These
data are the average of the AM and
PM response.
70
80
90