Apogee Instruments MO-200 소유자 매뉴얼 - 페이지 13
{카테고리_이름} Apogee Instruments MO-200에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Apogee Instruments MO-200 20 페이지. Oxygen meter
For measurements in soil or saturated air (100 % relative humidity), it is recommended that the MO-200 is calibrated in
conditions where relative humidity is 100 %. A simple way to accomplish this is to mount the sensor in a sealed chamber over
water, with ambient air filling the headspace, as shown below.
A) Relative humidity effects on relative
oxygen concentration shown as a function
of relative humidity at temperatures
increments of 10 C and B) as a function of
temperature at relative humidity increments
of 20 %. The air in soil is typically always
saturated with water vapor (100 % relative
humidity) unless the soil is very dry.
13