Apogee Instruments MO-200 소유자 매뉴얼 - 페이지 16
{카테고리_이름} Apogee Instruments MO-200에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Apogee Instruments MO-200 20 페이지. Oxygen meter
16
TROUBLESHOOTING AND CUSTOMER SUPPORT
Verify Functionality
Pressing the power button should activate the LCD and provide a real-time reading of percent oxygen concentration
(perform a calibration in ambient air if necessary). Since exhaled breath has a lower oxygen concentration than ambient air,
gently blowing across the sensor opening should force the displayed reading to a lower oxygen percentage. The reading will
take a few seconds to respond based on the sensor's response time. Move the sensor to a well-ventilated area and the
reading should return to the original reading (e.g., 20.9 %).
Battery Life
When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after continuous
use. The low battery indicator will appear in the upper left hand corner of the LCD display when the battery voltage drops
below 2.8 V DC. The meter will still function correctly for some time, but once the battery is drained the pushbuttons will no
longer respond and any logged measurements will be lost.
Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight amount of
current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is recommended to remove
the battery when storing the meter for many months at a time, in order to preserve battery life.
Low-Battery Error after Battery Replacement
A master reset will usually correct this error, please see the master reset section for details and cautions. If a master reset does
not remove the low battery indicator, please double check that the voltage of your new batter is above 2.8 V, this is the
threshold for the indicator to turn on.
Master Reset
If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after replacing the
old battery, a master reset can be performed that may correct the problem. Note that a master reset will erase all logged
measurements from memory.
First press the power button so that the LCD display is activated. While still powered, slide the battery out of the holder,
which will cause the LCD display to fade out. After a few seconds, slide the battery back into the holder. The LCD display will
flash all segments and then show a revision number (e.g. "R1.0"). This indicates the master reset was performed and the
display should return to normal.
Error Codes and Fixes
Error codes will appear in place of the real-time reading on the LCD display and will continue to flash until the problem is
corrected. Contact Apogee if the following fixes do not rectify the problem.
Err 1: battery voltage out of range. Fix: replace CR2320 battery and perform master reset.
Err 2: sensor voltage out of range. Fix: perform master reset.
Err 3 / Err 6: not calibrated. Fix: perform master reset. If reset doesn't work and the meter is nearing five years old it may
indicate the end of life for the oxygen sensor.
Err 4: CPU voltage below minimum. Fix: replace CR2320 battery and perform master reset.
Modifying Cable Length
Although it is possible to splice additional cable to the separate sensor of the MO-200, note that the cable wires are soldered
directly into the circuit board of the meter. Care should be taken to remove the back panel of the meter in order to access the