Aponorm BASIS PLUS BLUETOOTH 사용 설명서 지침 - 페이지 13

{카테고리_이름} Aponorm BASIS PLUS BLUETOOTH에 대한 사용 설명서 지침을 온라인으로 검색하거나 PDF를 다운로드하세요. Aponorm BASIS PLUS BLUETOOTH 19 페이지.

Error
Designa-
tion
«ERR 2»
Interference
signal
-C
17
«ERR 3»
Irregular
cuff pres-
-D
17
sure
«ERR 5» Irregular
measuring
result
Potential cause and
remedy
During the measurement,
interference signals
were detected at the
cuff, caused for instance
by movement or muscle
tension. Repeat the
measurement, keeping
your arm still.
It is not possible to build
up sufficient pressure in
the cuff. A leak may have
occurred. Check that the
cuff connection is correct
and that the cuff is not
too loose around your
arm. Replace the batter-
ies if necessary. Repeat
the measurement.
The measuring signals
are inaccurate and no
result can therefore be
displayed. Read through
the check list for perfor-
ming reliable measure-
ments and repeat the
measurement.*
Error
Designa-
tion
«ERR 6» MAM mode Too many errors occurred
«HI»
Pulse or cuff
pressure too
high
«LO»
Pulse too
low
* Please consult your doctor immediately, if this or
any other problem occurs repeatedly.
Potential cause and
remedy
during the measurement
in the MAM mode so
that no final result can
be determined. Read
through the check list
for performing reliable
measurements and then
repeat the measure-
ments.*
The pressure in the cuff
is too high (over 299
mmHg) or your pulse
rate is too high (over 200
beats per minute). Relax
for 5 minutes and repeat
the measurement.*
The pulse rate is too
low (less than 40 beats
per minute). Repeat the
measurement.*
13