Shodex RI-101
Response Time [s]
Sets the frequency of the data being sent to Clarity from the detector.
Setting this parameter to other value than Sampling Period means that the
detector will filter the data and only send averaged values to Clarity.
Temperature [°C]
Defines the working temperature of the Shodex RI-101 detector.
Equilibration Temperature Tolerance [°C]
Serves for the setting of the target temperature tolerance. When the
temperature reaches the desired value with the tolerance set here, the
detector gets to the READY state.
Autozero before Run
Sets whether the detector should be autozeroed at the beginning of the
analysis run.
From Det
Loads the detector control parameters from the detector to Clarity.
Det Status
When invoked, opens the
information regarding the connected detector.
Hardware Configuration
- 8 -
4 Using the control module
dialog showing the