DataApex Shimadzu LC-10 매뉴얼 - 페이지 14
{카테고리_이름} DataApex Shimadzu LC-10에 대한 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. DataApex Shimadzu LC-10 19 페이지. Metal free pump
4 Using the control module
4.1.1 Gradient Options
Invoke the Options... button in the
open the
name for particular solvents, to switch whether they are used or not in the
gradient and to set the warning levels for pressure to prevent the damage
to hardware.
Min. Pressure
Sets the minimum pressure for the given pump. When pressure drops to
the set value, the pump will shut down. This prevents the solvent leakage.
Max. Pressure
Sets the maximum pressure for the given pump. When pressure reaches
the set value, the pump will shut down. This serves to prevent the damage
to the pump when the column is blocked.
Solvent 1 (..4)
It is possible to enable/disable particular solvent, as well as to set custom
name to it.
Gradient Options
Fig 6: Gradient Options
Method Setup - LC Gradient
dialog. This dialog allows to set the custom
- 10 -
Clarity Controls
dialog to