Cleaning

Prior to cleaning, always disconnect the scale from the power supply. Do not use aggressive cleaning
agents to clean the scale; instead, use only a moist cloth or common disinfectants.
Make sure that no liquid enters the body of the scale or the display, since this can damage
the scale and make it unusable!

Error Messages

-
Low Battery
The scales alkaline AA type batteries are flat.
Replace the batteries with new ones
-
Over Load
This indicates that the scales load sensor/s
have been overloaded, re-calibration may fix
this problem but it usually indicates a faulty
load sensor
-
High/Low Zero Count
1. The scale is above its Zero Range, remove
any weight on the scales and try to power on
again. If the scale still shows the error
message then it indicates a fault with the
scales electronics.
2. The scale is below its Zero range, check
there is nothing jammed underneath the
scale and try to power on again, if the scale
still shows the error then it indicates a fault
with the scales electronics

Recalibration

Calibration is to be carried out only by authorized companies or persons in accordance with national
legislative regulations. The notified body (0122) and the year of initial calibration can be found on the
CE symbol. The scale must be recalibrated if
-
the calibration mark has been damaged
-
after a calibrated scale has been repaired
-
or after expiry of the period for recalibration stipulated by national statutory regulations.
M101000-01-210323-Rev005-UM-e

Care, servicing and recalibration