WARNING
Always install an intake filter, especially if the pump is to work with air that might contain foreign
matters.
Pump outlet gases must be treated in such a way that they do not contaminate the working environment and the
surrounding atmosphere. If condensable vapours are pumped in, the discharge hose should be routed downwards and
have no bends to prevent the condensate produced in the discharge line from building up or flowing back into the pump.
Never use friction hoses with hose diameters smaller than the intake diameter. Avoid exceeding hose lengths, tight
bends or bends spaced too closely together.
WARNING
Do not insert outlet fitting pipes or devices that block or impede the disposal of outlet gases (max
overpressure exhaust 0.3 bar).
4.9.3

WIRING

WARNING
Check that network voltage and frequency correspond to values contained on the motor rating plate.
The connection cable must be adequate for the power absorbed by the pump (absorption values are shown on the
pump motor rating plate) taking into account the environmental conditions of operation.
HAZARD
Always ground the pump.
Always install a security system between the pump and the electric power supply. Pump absorption values are
shown on the motor rating plate.
The pump is normally supplied without an electrical cable and switch. For electrical connection, see the diagram
contained within the terminal board or on the motor rating plate.
WARNING
Check that the direction of rotation is correct before starting the pump for the first time or after
changing the electrical connections.
The correct direction of rotation is indicated by the arrow on the pump (see ch. 3.3). Pump operation
with a rotation direction that is opposite to that indicated can severely damage the pump itself.
4.9.4

ELECTRICAL OIL LEVEL GAUGE

The electrical oil level gauge signals when the oil level inside the tank rises above the maximum level signalled.
Oil level gauge characteristics are as follows:
• VOLTAGE SWITCHING: Max 250Vdc & Vac
• CURRENT SWITCHING: Max 0.5 Adc / Max 0.7 Aac
• CONTACT RATING: Max 50VA
WARNING
The signal from this device may NOT be used to directly drive the motor.
8702037 – 01/01/2011 – R.0
Operating and Maintenance Instructions
LB.5WR – LC.25WR – LB.40WR – LB.60WR – LC.105WR