{카테고리_이름} Adec Decade 1021에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. Adec Decade 1021 14 페이지. Vac back chair

2. Is the manual water shutoff valve in the
floor box fully OPEN (refer to your Floor
Boxes Owner's Guide, Publication No.
85-2611-00)?
Yes: Proceed with Step 3.
No: Turn the valve fully counterclockwise,
then retest the syringe.
3. Remove and inspect the water filter screen
(refer to your Floor Boxes Owner's Guide,
Publication No. 85-2611-00). Is the screen
clogged?
Yes: Replace the filter and retest the syringe.
No: Look for pinched or crimped tubing
between the floor box and the syringe.
4. Is the syringe water flow pinch valve
completely closed?
Yes: Turn the water flow adjustment screw
counterclockwise to the desired flow,
then retest the unit.
No: Look for pinched or crimped tubing
between the floor box and the syringe.
Inadequate or No Vacuum
1. Is the central vacuum turned on?
Yes: Proceed with Step 2.
No: Turn on the central vacuum and retest.
2. Remove and inspect the solids collector
screen (refer to page 4). Is the screen clogged
with debris?
Yes: Clean the screen and retest.
No: The vacuum supply to the Decade 1021
Vac Back Chair is clogged. Contact your
authorized A-dec dealer for assistance.
If the preceding steps failed to correct the
problem, contact your authorized A-dec dealer
for further recommendations.
Decade 1021 Vac Back Chair