2.1. MODE OF USE

The method of application of the device is intermittent. The operational cycle indicated respects the following times:
Type of Instrument
i-MMs
i-MMr
i-MMr L

2.2. ENVIRONMENTAL CONDITIONS FOR USE

• Ambient Temperature 10 ÷ 40 °C
• Relative Humidity 30 ÷ 75 %
• Atmospheric Pressure 700 ÷ 1060 hPa (700 ÷ 1060 mBar)

2.3. ENVIRONMENTAL CONDITIONS PERMITTED FOR TRANSPORT AND WAREHOUSING

• Ambient Temperature between -20 ÷ +70 °C
• Relative Humidity 10 ÷ 100 %
• Atmospheric Pressure 500 ÷ 1060 hPa (500 ÷ 1060 mBar)

2.4. APPLIED PARTS

The parts that, during standard use, necessarily come into contact with the patient in order for the device to carry out its functions correctly, are the
handle covers.
3.

OPERATION OF THE DEVICE

3.1. CONNECTION TO THE POWER SUPPLY INSTRUMENT TUBING

Bring the motor next to the instrument tubing and rotate it, on the connector itself until the coupling point is found. Insert the motor completely and then
fully tighten the ring nut.

3.2. CONNECTION AND DISCONNECTION OF THE HANDPIECE

Handpieces complying with ISO 3964 and with the INTRAmatic
is heard.
Never mount a handpiece onto a motor that is running.
To detach the handpiece from the motor pull lightly.

3.3. OPERATING PARAMETERS

Operational modes, speed adjustment, inversion of the rotation direction, continous or reciprocating movements, torque adjustment, spray and LED
lighting activation (if available) are to be made through the Dental Unit command panel. Refer to the Dental Unit instruction manual for the setting of the
operational parameters.

3.4. ILLUMINATION DEVICE

In the micro-motors set up with LEDs for lighting of the surgical field, these light up automatically when the micro-motor is running.
After the motor stops, the LED remains lighted for a pre-set time. Lighting depends on settings made on the Dental Unit.
EN
6
I-MMr • i-MMr L • i-MMs
Operation (minutes)
5
5
5
Lux. type may be used. Insert the handpiece on the coupling until the connection "click"
®
Pause (minutes)
5
5
5