HP- 30/40/50
CAUTION
! 
●Be sure to unplug the pump unit. 
●Replace the diaphragms and the valves with new ones at least once a year to one and 
a half year regularly in order to maintain their initial performance. 
●For  chamber  block  replacement,  be  sure  to  change  both  chamber  blocks  at  the  same 
time. 
●The  rod  employs  powerful  permanent  magnets.  Therefore,  be  sure  to  remove  your 
watch and any other precision machines before operation as they may be affected by 
the strong magnetic force. 
●Do  not  put  the  actuating  rod  close  to  a  magnetic  card,  a  magnetic  disk  or  other 
magnetic media as the data may be destroyed.
STEP
STEP
REMOVAL OF THE CHAMBER BLOCKS
1
1
To remove the upper housing. 
To remove the sound absorber. 
Pull out the L-tube from the casing block. 
Remove  the  4  screws  holding  the  chamber  block 
and the casing block. (4 screws on each side) 
 
STEP
STEP
2
2
Remove 1 U-lock nuts from one side holding the 
diaphragm mounting block to the rod. 
●Use the nut driver to loosen (or tighten) the U-lock nut.
STEP
STEP
3
3
Remove  one  of  the  diaphragm  mounting  blocks 
from  the  actuating  rod  and  pull  out  the  other 
diaphragm  mounting  block  with  the  rod.  After 
that, separate the diaphragm mounting block and 
the rod. 
This  completes  the  chamber  block  removal 
procedure.
2
REPLACING THE CHAMBER BLOCK