activecare medical medalist 소유자 매뉴얼 - 페이지 16
{카테고리_이름} activecare medical medalist에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. activecare medical medalist 20 페이지.
For service assistance, please call your authorized ActiveCare dealer.
The diagnostic flash codes for your power wheelchair are designed to help you perform basic
troubleshooting quickly and easily. A diagnostic flash code flashes from all 6 of the LED's
located on the controller in the event one of the conditions listed below develops.
Table 1: Power Wheelchair Controller Internal Diagnostics
Flash
Code
1
Motor Current Over Limit
2
3
4
5
Left Parking Brake
6
Right Parking Brake
7
8
9
If more than one fault exists, the fault having the highest priority (and lowest number of
flashes) is indicated. All faults disable the controller and require the controller to be turned off
then on again once the source of the fault is removed.
15
Troubleshooting
Description
Battery
Left Motor
Right Motor
User Interface
Controller
A-series Bus
Communication
Recommended Action
Motor current is over the controller set limit.
Release the joystick to neutral and try again
Try charging the batteries.
Batteries need replacing.
Check the battery cables and connections
Check left motor cables and connections
Check right motor cables and connections
Check left parking brake cables and connections
Check right parking brake cables and
connections
Display PCB fault. Consult a service agent.
Consult a service agent.
Consult a service agent.
Medalist Owner's Manual