Gait up Gait Up 사용자 설명서 - 페이지 4

{카테고리_이름} Gait up Gait Up에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Gait up Gait Up 17 페이지.

Gait Up SA
EPFL Innovation Park, Bat C. CH-1015 Lausanne
www.gaitup.com
TVA CHE-145.136.760

1. GaitUp Lab package content

2. Get ready with the sensors

2.1 Charging sensors

®
Physilog
5 are charged via micro USB. Connect the USB cable delivered with the sensors.
Connect the other part to a computer or sector with an adaptor (voltage specification: 5V,
5 must be turned off when charging. Don't charge the Physilog
®
®
100 mA). Physilog
5 if the
temperature is not between 0 ° and 45 °C.
Low battery level is indicated by orange blinking during measurement or by 3 rapid red blinks
®
when you want to start a new measurement. If the Physilog
5 is completely discharged,
charge at least 30 minutes before using. Battery level estimation is available when you pair
the sensor to the GaitUp application (available on Apple Store and Google Play Store).
NOTE: Charging problem is indicated by a continuous red light and it could damage
the sensor. In this case try to charge your sensor on another charging source.

2.2 Sensors Configuration

By default, sensors are configured for running analysis with the GaitUp Lab. If you encounter
any problem during the analysis you can check the sensor configuration with the Research
Toolkit (Windows version/Mac version, User guide) in the "Configuration" part. Select your
sensors> "Load Config"> Change the configuration> "Send Config". For walking analysis,
the same sensor configuration as for running can be used. But if you only use walking
analysis, it is possible to set the sensor configuration to a lower frequency to reduce the size
of the raw data files (see table below) using the Research Toolkit.
User guide GaitUp Lab - v1.0.0 Date: 04.05.2020
Copyright © 2020, All Rights Reserved - Gait Up SA. CH-1015 Switzerland
4