OPERATING PROCEDURE

1. Locate and position the LifeChoice Activox POC so that the Fan Inlet
Vent and Fan Outlet Vent are not obstructed. The unit, particularly
the air inlet slots and the air outlet areas, should be positioned in a
well-ventilated space.
3. Press the Power Button once to turn the unit on. An audible beep will
NOTE: After the initial power on, the Pulse Setting LED corresponding
to the last flow rate used will illuminate.
CAUTION: DO NOT place the cannula into nasal passages until the
!
LifeChoice Activox POC is fully powered on. Breathing through the
cannula during start-up will prevent the system from powering on.
12
2. Power the LifeChoice Activox POC
using one of the four modes described
in the Powering the LifeChoice
Activox POC section above.
occur and the Power LED, the Auto
Mode Active-High LED and the Pulse
Setting of 1 LED will illuminate solid
green. This indicates that the Life-
Choice Activox POC is powered on,
is ready for use, and that settings
may now be changed. Pressing the
Power Button for an extended time
after the audible beep will cause the
system to turn back off.
4. There are two inhalation sensitivity
modes on the control panel:
Active-High and Sleep-Low which
automatically adjust based on oxygen
demand (e.g. breathing pattern)
requirements. No action is required
to start the delivery of oxygen.
5. Select a Pulse Setting of 1, 2, or 3 as prescribed by your doctor.
These settings are equivalent to 1, 2, or 3 LPM settings on a
continuous flow device. The green LED above the selected setting
illuminates and will flash each time inhalation is detected indicating
that a pulse-wave of oxygen has been delivered.
6. Place the nasal cannula in the nasal passages and breathe normally.
When the LifeChoice Activox POC senses inhalation, oxygen is
supplied through the cannula. The duration and size of the pulse
is dependent upon the Pulse Setting selected.
7. Check to make sure your unit's
battery is fully charged before
venturing out with the LifeChoice
Activox POC for the first time. (See
Battery Charge Indicator section,
User Controls).
8. To turn the LifeChoice Activox POC
off, press the Power Button and
hold for 2 seconds until the green
Power LED goes off and the unit
shuts down. After the LifeChoice
Activox POC is powered off, the
internal fans may continue to
operate for an additional period of
time until the temperature inside
the unit is sufficiently cool.
13