iHealth BPM1AE 소유자 매뉴얼 - 페이지 10

{카테고리_이름} iHealth BPM1AE에 대한 소유자 매뉴얼을 온라인으로 검색하거나 PDF를 다운로드하세요. iHealth BPM1AE 17 페이지.

The monitor, cable, battery and cuff must be disposed of
according to local regulations at the end of their usage.
Note
The battery has limited charge cycles and may eventually
need to replaced by an iHealth service provider. Battery life
and charge cycles vary by use and settings.
Monitor
Status
Charging
Fully
charged
Low battery

TECHNICAL ALARM DESCRIPTION

The monitor will show 'SYS or DIA beyond measurement range' or
' SYS or DIA below measurement range'' as technical alarm on screen
with no delay if the determined blood pressure (systolic or
diastolic) is outside the rated range specified in part
SPECIFICACIONS. In this case, you should consult a physician or
check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset
in the factory and cannot be adjusted or inactivated. This alarm
condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal
displayed on screen will disappear automatically after about 8
seconds.

TROUBLESHOOTING

PROBLEM
Low Battery
Display reads
"ERR xxx"
Status Indicator
dynamical battery symbol
full battery symbol
empty battery symbol and reminder"Low battery
unable to perform measurement, please
recharge"
POSSIBLE CAUSE
Battery is less
than 20%
Blood pressure is
outside of
measurement range
Arm or monitor was
moved during test
The cuff does not
inflate properly
or pressure falls
quickly during
test
SOLUTION
Charge the battery
Retest, make sure your
blood pressure is within
measurement range
Retest, make sure not to
move your arm or the
monitor
Review the cuff
applizcation instructions
and retest
10