iHealth PT3SBT 사용자 설명서 - 페이지 4
{카테고리_이름} iHealth PT3SBT에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. iHealth PT3SBT 18 페이지. Wireless no-touch forehead thermometer
such tissues will affect temperature conduction of the body.
7. If there is a temperature difference between the thermometer storage area
and the new ambient environment around the subject, please let the
thermometer sit within the new environment for 30 minutes before taking the
measurement.
8. Do not measure body temperature immediately after consuming a drug that
raises body temperature. Temperature readings taken at this time will not be
accurate.
9. It is normal for readings taken from continuous measurements to fluctuate
within a small range. During continuous measurements, the subject's body
temperature may be transmitted to the thermometer, affecting measurement
accuracy. We recommend taking only up to 3 continuous readings within a
short period.
10. Do not directly face the sun or an air outlet of an air conditioning or radiator
device during the measurement as this will cause changes to the forehead
temperature. Measurements should be taken in a stable environment where
possible.
11. Do not measure body temperature in an environment with strong EM
interference (examples include places close to a working microwave, induction
cooker, or cellphone in-use) as EM interference may cause errors in the
reading or even device failure.
12. Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of the product. Otherwise, degradation of the
performance of this equipment could result.
13. This product should be considered a personal device. Clean and sanitize
the product properly to prevent cross contamination.
14. To clean a dirty thermometer probe, gently swipe the probe using a cotton
swab dipped in 70% isopropyl. Let the cleaned thermometer sit for at least 15
minutes before taking more measurements.
15. This infrared thermometer meets requirements established in ISO
80601-2-56:2017/AC:2018. It displays subject's temperature over a range of
89.6℉~109.4℉(32℃~42.9℃). Full responsibility for the conformance of this
product to the standard is assumed by (manufacturer: ANDON HEALTH CO.,
LTD. Add: No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190,
China).
(2) About the product
1. This product is a precision device. Return the product to its original
packaging for proper storage after use. To ensure accurate measurements,
avoid the device or probe contacting any liquid or droplets. Avoid tiny particles
(such as dust or powder) falling into the probe.
2. Avoid dropping or subjecting the product to external forces. Do not
disassemble or re-assemble the product on your own.
4