SERIES 5 SIDE DELIVERY SYSTEM
Instructions For Use
MAINTENANCE
DENTAL UNIT - WEEKLY MAINTENANCE
A cleaning procedure should be performed at least once a week, preferably at the start of the week before treating
patients. If the unit is to be stored for any length of time, perform a weekly maintenance routine immediately before and
after storage.
1.
Purge the unit with air.
2.
Flush the system with disinfectant solution:
a. Turn the unit Master switch to the Off position. Empty the water bottle, replacing the water with cleaning
b. Hold the handpiece tubing and syringe over the cuspidor or other suitable container. Turn the unit on, wait
3.
Allow the disinfectant to remain in the unit for 10 to 20 minutes, then flush the system again until all the cleaning
solution is used up.
4.
Purge the unit with air:
a. Hold the handpiece tubing and syringe over a container. Turn the Master Switch to the On position, wait a
b. Turn the unit Master Switch to the Off position. (If the unit will be stored, stop here.)
5.
Fill with clean water:
a. With the unit Master Switch turned to the Off position, remove the empty disinfectant bottle. Replace with
b. Hold the handpiece tubing over a suitable container. Turn the unit on, wait a few moments, then operate the
HANDPIECE OIL COLLECTOR
Replace the 2" x 2" gauze pads with clean pads in the
handpiece oil collector at least every 90 days, or more
often if handpieces are oiled frequently. Remove the oil
collector by squeezing the sides and pulling downward.
16
solution.
a few moments, then operate the flush toggle, syringe and foot control until a continuous stream of cleaning
solution is running through the system.
few moments, then operate the flush toggle, syringe and foot control until all cleaning solution is purged from
the system.
clean bottle and clean water.
flush toggle until a continuous stream of water is flowing through the system. Replace handpieces and do the
same with the syringe. The unit is now ready for use.
OIL COLLECTOR
16