Actonis DUAL ACTION 18 사용자 설명서 - 페이지 11

{카테고리_이름} Actonis DUAL ACTION 18에 대한 사용자 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Actonis DUAL ACTION 18 17 페이지.
Actonis DUAL ACTION 18에 대해서도 마찬가지입니다: 빠른 시작 매뉴얼 (2 페이지)

ELECTROMAGNETIC IMMUNITY GUIDANCE

Guidance and manufacturer's declaration – electromagnetic immunity
The LaserComb is intended for use in the electromagnetic environment specified
below. The customer or the user of the LaserComb should assure that it is used in
such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
LUH018DWEDXXX-00-0723.indd 11
LUH018DWEDXXX-00-0723.indd 11
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2kV for power supply lines
± 1kV for input/output lines
± 1kV line(s) to line(s)
± 2kV line(s) to earth
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2kV for power supply
lines
Non applicable
± 1kV differential mode
Non applicable
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 sec
3 A/m
Electromagnetic
environment-
guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
LaserComb require
continued operation
during power mains
interruptions, it is
recommended that
the LaserComb be
powered from an
uninterruptible power
supply or a battery.
The LaserComb power
frequency magnetic
fields should be at
levels characteristic of
a typical location in a
typical commercial or
hospital environment.
10
7/18/23 1:53 PM
7/18/23 1:53 PM