Water Control Knob: For adjusting the flow of water to the
handpiece and insert tip. To obtain the necessary water flow of at
least (18cc - 30cc/+/minute) turn the WATER KNOB clockwise.
HIGHER water flow to the RIGHT, and LOWER water flow to the
LEFT.
Handpiece Panel Receptacle: Accepts the male side of the
Handpiece connector with lock ring.

HANDPIECE ASSEMBLY

The handpiece assembly is used to accept Cavitron® interchangeable
insert tips. It is connected to the AUTOSCALER® front panel by a Quick
Connector. The open end of the handpiece will accept all Cavitron®
brand: "P", "TFI™", "SLIMLINE™" or "FSI™" style 25,000Hz (25K)
inserts. If you ordered a 30,000Hz (30K) model AUTOSCALER®, you
must use the shorter (30K) frequency insert tips.
30K tips measure (Shorter style approx. 6.25 inches long)
25K tips measure (Longer style approx. 7.25 inches long)
(25K) inserts will NOT work in your 30K unit.
(30K) inserts will NOT work in your 25K unit.
Inserts are extremely FRAGILE. If one is dropped, bent, or scratched it
WILL NOT cavitate (i.e. will not function) or function by operating at a
higher than normal temperature, and should be replaced immediately.
Insert tips should be routinely replaced according to insert
manufacturer's recommendations.

ELECTRICAL REQUIREMENTS

A GROUNDED wall receptacle is required supplying a voltage range
of 110-125 volts 50/60 Hz A.C. If a 3 to 2 prong adapter is used, the
ground wire MUST be connected to the base plate screw. Have an
electrician test your outlet for correct wiring polarity and a safe
ground. Do not assume the ground is complete. If you plug the
AUTOSCALER® in a GFI (Ground Fault Interrupt ) outlet it may trip the
socket due to the high sensitivity of the socket. If this occurs plug the
AUTOSCALER® into a non-GFI socket.
*Cavitron®, SLIMLINE™, FSI™, and TFI™ are registered trade marks of Dentsply Sirona
7

INSTALLING THE HANDPIECE ASSEMBLY

1. Align the electrical contacts of the MALE connector with the same on
the FEMALE panel connector.
2. Gently insert the connector. It should slide in with very little effort. If
the connector does not align, DO NOT force it. Turn the CONNECTOR
BODY in either direction until the connector correctly aligns and goes
in.
3. Once the connector has been completely inserted, screw the OUTSIDE
LOCK RING to the RIGHT with light finger pressure until it stops. DO
NOT over tighten the lock ring, be careful not to cross thread the lock
ring.
4. To remove the connector, unscrew the lock ring to the left, and gently
pull outward grasping the entire connector body.

INSTALLATION AND OPERATION

DO NOT use your AUTOSCALER® until all of the following steps have been
completed.
1. Upon receipt of your unit carefully unpack it and see that the Toggle
Switch is in the OFF-O position.
2. The Water Control Knob should be in the HIGH position by turning
the water control knob all the way to the RIGHT.
3. Set the Power Knob all the way to the LEFT in the LOW position.
4. Plug the male Water Quick Release Valve water connector into the
Dental operatory or have a plumber install a SHUT OFF style FEMALE
connector. A minimum of (20 P.S.I.) water input pressure and a
maximum of (60 P.S.I.) is needed from the operatory for proper oper-
ation of the AUTOSCALER®. It is recommended that water line filtra-
tion be used to remove larger particles before entering the scaler.
5. Plug the power cord into the wall (see electrical requirements). Press
the toggle switch to the ON-I position. The L.E.D. light above the
switch should be ON ("I" position).
6. Select an insert. Moisten the rubber O-RING before inserting into
the handpiece. While holding the handpiece UPRIGHT place the
insert in the open end of the handpiece and with a gentle PRESS and
TURN motion, snap the insert in fully. DO NOT force the insert.
7. Manufacturer recommends the use of a disposable handpiece
sleeve/barrier be used over the handpiece assembly. Place the
sleeve/barrier over the handpiece. While holding the sleeve/
barrier in place, puncture the insert through the plastic at the open
end of the handpiece, and snap the insert in fully. Dispose of
sleeve/barrier after each patient.
8. Due to atmospheric exposure and exposure to sterilization solutions
the O-RING on the insert tip may become swollen, brittle, or crack,
and require replacement.
8