Cell ushing system guide
The membrane cell is sensitive and requires regular checks and maintenance. The cell will
require descaling to remove minerals and limescale deposits.
This is done by ushing the cell through with a weak acid solution. It is recommended to
use either white vinegar or a diluted hydrochloric acid solution.
To undertake the cell ushing process:
Before following this process, activate the
'E-STOP' button on the front of the Flow device
Do not plug the Flow wash system into the Flow
device until instructed.
Remove the pipe from the Catholyte output.
Push the push- t ring around the tube into the
device as shown
Whilst holding the circular ring into the device,
pull the tube out of the xing simultaneously.
The pipe will detach from the tting.
CENTREGO FLOW 40
Remove Anolyte pipe from device
To remove the Anolyte pipe, nd the 'T' shaped
connector as shown here with a yellow arrow.
This is inside the device at the bottom right-hand side.
Use the push- t system to disconnect the pipe.
The image shows the view after the pipe
has been disconnected.
From the outside of the Flow 40 device, the pipe will
now be loose but still be inserted. You do not need
to rotate the black screw- xing to remove
the pipe.
Lightly pull the pipe out of the Anolyte connection.
CENTREGO FLOW 40
P 5