Major tech MT715 사용 설명서 - 페이지 12
{카테고리_이름} Major tech MT715에 대한 사용 설명서을 온라인으로 검색하거나 PDF를 다운로드하세요. Major tech MT715 16 페이지. Ac trms open jaw clamp meter
5.6. Capacitance Measurements
WARNING: To avoid electric shock, discharge the capacitor under test
before measuring.
1.
Set the function switch to the Ω
2.
lnsert the black test lead banana plug into the negative COM jack and
the red test lead banana plug into the V Ω
3.
Use the MODE button to select CAP Measurements.
4.
Touch the test probe tips across the part under test.
5.
Read the capacitance value in the display.
6.
The display will indicate the proper decimal point and value.
For very large values of capacitance measurement it can take several
minutes before the final reading stabilizes.
5.7. Diode Measurements
1. lnsert the black test lead banana plug into the negative COM jack and
the red test lead banana plug into the V Ω
2. Turn the function switch to Ω
button to select the diode function if necessary (diode symbol will
appear on the LCD when in Diode test mode).
3. Touch the test probe tips to the diode or semiconductor junction under
test. Note the meter reading.
4. Reverse the test lead polarity by reversing the red and black leads.
Note this reading.
5. The diode or junction can be evaluated as follows:
•
If one reading displays a value (typically 0.400V to 0.900V) and the
other reading displays "0L" ,the diode is good.
•
If both readings display "0L" the device is open.
•
If both readings are very small or "0", the device is shorted.
5.8. Non-Contact Voltage (NCV) Measurements
The NCV function works on any rotary switch position.
1. Test the detector on a known live circuit before use.
2. Hold the top of the meter very close to the voltage source as shown.
3. If voltage is present. the red light will long lighting.
NOTE: Do not touch the top of the meter when using this function.
NOTE: Test on known live circuit before using.
12
CAP position.
CAP positive jack.
CAP positive jack.
CAP position. Use the MODE