5. Operation

5.1 Preparing the test
• Prior to every test, please ensure that the instrument is in perfect
condition:
• For example, keep an eye out for a broken housing or leaking batteries.
• Always carry out a function test before using the voltage tester, see
below.
• Check that the instrument is functioning properly(for example at a known
voltage source)before and after every test.
• If the safety of the user can not be guaranteed, switch off the instrument
and secure it to prevent unintentional usage.
Carrying out a function test
•When press "Autotest" button, with the exception of the LED for low
impedance test, all LEDs should light up. If the battery voltage is too low,
all leds will blink, please replace the battery
5.2 Preparing the test
• Connect both test probes with power source.
• As from a voltage of > 6V the voltage tester
switches on automatically.
• The voltage is displayed via LEDs.
• The different indicating signals of the voltage
detector (including the ELV lim indication)are not
to be used for measuring purposes.
• For AC voltage, the "AC" is illuminated.
• For positive voltage, the "+" is illuminated.
• For negative voltage, the "-" is illuminated.
• In the case of DC voltage, the polarity of the
indicated voltage relates to the voltage tester probe tip
• Once the safety extra-low voltage (50V AC /120V DC) is reached or
exceeded, the "
main circuit failure, an acoustic signal is emitted.
" is illuminated, in the event of no battery power or
10