Tube placement micro rotors
ALWAYS
load your tubes opposite to each other
ALWAYS
weigh samples to within 0.1 Gram. For best separations. (see page 4)
ALWAYS
fit the lid supplied
CHECK
the Rim seal gasket and replace every 2 months (Part number 8011)
CHECK the rotor screw fixing tightness regularly
...........................................................................................................................................................................
ALWAYS
load your tubes opposite to each other
ALWAYS
weigh samples to within 0.1 Gram For best separations. (seep page 4)
ALWAYS
fit the lid supplied
CHECK
the two top seals and replace regularly (Part number 8044)
CHECK the rotor screw fixing tightness regularly
Haematocrit Rotor
Always fit the lid supplied, as shown to the left
Micro tube Rotor
Always fit the lid supplied, as shown to the left